LF-16U Series U-Type Mass Flow Meter

The LF-16 series mass flow meter is a high-precision measuring device capable of directly detecting the mass flow rate of fluids—whether liquids, gases, or multiphase flows. Its key advantage lies in its ability to provide accurate mass flow data without requiring compensation for parameters such as temperature or pressure, effectively eliminating measurement errors caused by fluid density variations that commonly occur with traditional volumetric flow meters. This product features the following characteristics:
● Measures gas mass flow without requiring temperature or pressure compensation
● Large range ratio, measuring flow velocity range: 0.1 Nm/s to 120 Nm/s. ■ No pressure loss, suitable for pipes of any shape with known cross-sectional areas.
● Corrosion-resistant sensors, ideal for measuring corrosive gases. The insertion-type sensor allows for convenient online installation and maintenance.
● The expert algorithm covering the full measurement range ensures high accuracy, making it suitable for trade settlement or gas leak detection.

Key words:

Classification:

Download information:

Product Description

Product Overview Product overview

The LF-16 series mass flow meter is a high-precision measuring device capable of directly detecting the mass flow rate of fluids—whether liquids, gases, or multiphase flows. Its key advantage lies in its ability to provide accurate mass flow data without requiring compensation for parameters such as temperature or pressure, effectively eliminating measurement errors caused by fluid density variations that often plague conventional volumetric flow meters.

Performance Metrics TECHNICAL PARAMETERS

Measuring the density range

0~2 g/cm³

Accuracy and Precision

0.1~0.2

Repetitiveness

0.1–0.25%

Converter format

Wall-mounted type, on-site type

Nominal Diameter

DN6-DN200

Measuring medium

In addition to uniformly flowing fluids with normal viscosity, it can also measure high-viscosity fluids (such as honey, oil pastes, etc.); moreover, it is not limited to measuring just a single

The fluid parameters of the solution can also measure multiphase flows.

Medium temperature

-50℃ to 200℃

Material

Measurement tube made of 316L stainless steel, outer casing made of 304 stainless steel

Work stress

0–4 MPa

Power Supply

220V AC ±10%, 50Hz or 24V DC ±5%

Explosion-proof rating

Exd (ia)ⅡCT6GB

Signal Output

Pulse output, 4-20mA current output, and optional RS485 output

Measurements show

Instantaneous flow rate, cumulative flow rate, density, temperature

Sensor Protection Rating

IP65 (wall-mounted type), IP67 (field-type)

How it works   Working Principle

The principle behind mass measurement in Coriolis mass flowmeters is Newton's Second Law, F = Ma. When a fluid flows through the vibrating tube, it generates a Coriolis force that is directly proportional to the mass flow rate. If no fluid is flowing, the vibrating tube remains untwisted, and the electromagnetic signal detectors on either side of the tube pick up signals that are in phase. However, when fluid passes through, the vibrating tube twists under the influence of the torque, creating a phase difference between the two detectors. The transmitter measures the time lag between the left and right detection signals, and multiplying this time difference by the flow calibration coefficient allows it to determine the mass flow rate accurately.
The density measurement principle of a Coriolis mass flowmeter is based on the fact that the vibration frequency is inversely proportional to the square root of the fluid density—thus, by measuring the vibration frequency, the fluid density can be determined. Therefore, a mass flowmeter can simultaneously measure both the mass flow rate and the density of the fluid.

 

External dimensions APPEARANCE OF S

                                                                

Installation Requirements INSTALLATION

1. The flowmeter can be installed on horizontal, vertical, or inclined pipelines and positioned at any angle around the pipe circumference—provided ease of replacement and maintenance is considered. For dirty or moist gases, it is recommended to install the flowmeter horizontally (pointing upward) to minimize the risk of sensor contamination. Additionally, the sensor should be mounted away from the lowest point of the pipeline to prevent water accumulation inside the pipe, which could interfere with the proper operation of the flowmeter.
2. The straight pipe section requires 5 times the diameter (D) upstream and 3 times the diameter (D) downstream. However, when a gas rectifier is installed or the gas flow rate is below 30 m/s, the requirements can be reduced to 3 times D upstream and 2 times D downstream.
3. In explosion-proof areas, open flames or pipelines made of non-weldable materials are prohibited. Instead, dedicated clamps can be used to securely attach sensors to the pipes. In explosion-proof zones, use a manual drill to create holes, while in safe areas, an electric drilling tool is recommended.
4. The arrow on the sensor housing should point toward "1," aligning with the direction of gas flow. The sensor should be inserted perpendicular to the direction of gas flow.

Product inquiry

If you are interested in our products, please leave your e-mail, we will contact you as soon as possible, thank you!

Submit

Copyright©2022 Tianjin Ling Yu Technology Co., Ltd